Effect of Column Inlet and Outlet Geometry on Large-scale HPLC

Author(s)
Tan, S.N.; Khoo, Boo Cheong
Thumbnail
DownloadHPCES008.pdf (375.6Kb)
Metadata
Show full item record
Abstract
The separating characteristics of high performance liquid chromatography (HPLC) columns, measured in terms of the height equivalent of a theoretical plate (HETP) and skewness of the eluted peak, are investigated using computational fluid dynamics (CFD). Gradually expanding and contracting sections are introduced at the inlet and outlet, respectively, in columns with and without frits and their performance was compared with that of the conventional columns without expanding and contracting regions.
Date issued
2003-01
URI
http://hdl.handle.net/1721.1/3700
Series/Report no.
High Performance Computation for Engineered Systems (HPCES);
Keywords
column efficiency, computational fluid dynamics, FLUENT, Frit quality, liquid chromatography

Collections
  • High Performance Computation for Engineered Systems (HPCES)

Browse

All of DSpaceCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

Login

Statistics

OA StatisticsStatistics by CountryStatistics by Department
MIT Libraries
PrivacyPermissionsAccessibilityContact us
MIT
Content created by the MIT Libraries, CC BY-NC unless otherwise noted. Notify us about copyright concerns.